A Gas-based neutron imaging detector with individual read-outs

Nakamura, Tatsuya; Tanaka, Hiroki; Yamagishi, Hideshi; Soyama, Kazuhiko; Aizawa, Kazuya; Ochi, Atsuhiko*; Tanimori, Toru*

Nuclear Instruments and Methods in Physics Research A, 573(1-2), p.187 - 190, 2007/04

 Times Cited Count:9 Percentile:56.35(Instruments & Instrumentation)

We have been developing a neutron imaging gas detector with a high spatial resolution and with a high temporal response for the neutron scattering instruments at the pulsed neutron source in the Japan proton accelerator research complex. The gas detector system with individual read-outs was developed to meet the requirements for the instruments for neutron reflectometry or for small angle neutron scattering. The performances of the prototype detector using a multi-wire (MW) or micro-strip (MS) detector head were evaluated using a collimated neutron beam, and we confirmed the MS detector exhibiting a spatial resolution of 1.5 mm and a pulse-pair resolution of about 100 ns with a gas pressure of 6 atm helium with a mixture of 30% CF$$_{4}$$. The performances for the MS detector were also evaluated up to the total gas pressure of 8 atm.

JAEA Reports

Development of 2-d position-sensitive neutron detector with individual readout; Operation test and establishment of detection system by means of neutron beam

Tanaka, Hiroki; Yamagishi, Hideshi; Nakamura, Tatsuya; Soyama, Kazuhiko; Aizawa, Kazuya

JAERI-Research 2005-010, 16 Pages, 2005/04

JAERI-Research-2005-010.pdf:3.39MB

We have been developing the 2-d position-sensitive neutron detector with individual readout as next-generation-type detector system for neutron scattering experiments using intense pulsed neutron source. The detection system is designed to fulfill the specifications required for each neutron spectrometer, such as a count rate, efficiency, neutron/$$gamma$$-ray ratio, a spatial resolution and a size, by using suitable detector heads. The fundamental and imaging performances of the developed system assembled with a Multi-wire proportional counter head were evaluated using a collimated neutron beam. The system worked stably for long hours at the gas pressure of 5 atm with a mixture of 30% C$$_{2}$$H$$_{6}$$ (0.26 atm $$^{3}$$He)at gas gain of 450. The spatial resolutions were 1.4, 1.6 mm (FWHM) for a cathode- and a back strip- direction, respectively, considering a beam size. It was also confirmed that the spatial uniformity of the detection efficiency over the whole sensitive detection area was rather good, $$pm$$8 % deviation from the average with the optimum discrimination level.

Journal Articles

Development of micro-strip gas chamber with individual readout

Nakamura, Tatsuya; Masaoka, Sei; Yamagishi, Hideshi; Soyama, Kazuhiko; Aizawa, Kazuya

Proceedings of ICANS-XVI, Volume 1, p.441 - 446, 2003/07

Development of a Micro-Strip Gas Chamber(MSGC) in JAERI is presented. The performances we aim at are counting rate $$>$$10$$^{6}$$${it cps/mm}$$$^{2}$$, position resolution $$<$$1${it mm}$, detection efficiency $$>$$60% at thermal neutron and good n/gamma discrimination. Our two-dimensional gas detector consists of a micro strip plate printed on a polyimide substrate and a high pressure gas chamber which has all the signal lines are directly taken out (256ch individual readout) and withstands the pressure of 10bar $$^{3}$$${it He}$. The individual signal readout opens the way to realize high counting rate and high position resolution at the same time. We carried out range measurements of secondary particles to confirm the range measuring power of the MSGC with that read out method. The measured ranges in the gas of 3bar helium mixed with $$10%$$ ethane is 3mm, 9.5mm for triton and proton, respectively. Those values are well agreed with theoretical estimation and we could show the feasibility of the precise measurement of the ranges by the MSGC.